Advanced technology is used to rapidly measure and detect the surface particles of products.
Particle size channel
0.1um,0.2um,0.3um, 0.,5um. 3.0um, 5.0um
LPC LS-60
Origin
American LIGHTHOUSE
Product Features
Use liquid (pure water) sampler combined with optical particle technology to count particles;
Test Range
8 channels, 0.2 -2.0μm;
Test Methods
First test the pure water to get the particle number 1 as the background value, and then put the test sample into the pure water to test the particle number 2.
After subtracting the particle number 1 from the particle number 2 and dividing the sample area, the quotient is the number of particles per unit area (pieces/cm2) of the sample.
High intensity UV Lamp
Origin
U.S. UVP
Product Features
Use ultraviolet light to make particles with fluorescent properties reflect light for sample surface particle detection
Test Range
Darkroom operation, static test;
Test Methods
Place the UV light lamp on the test bench in the dark room, and place the sample under the lamp,
Press the power button, the UV light gradually brightens,
Particles containing fluorescent objects will show reflections.
HHPC3+
Origin
United States Metone
Product Features
Counting gas particles with laser particle sizer based on MIE laser scattering principle;
Test Range
3 channels, 0.3μm, 0.5μm, 1.0μm.;
Test Methods
Install the self-cleaner to the air inlet of the instrument, and self-clean for one minute;
Remove the self-cleaning device, test, the instrument will automatically stop after running for one minute,